6100B/6105A
Users Manual
5-20
Keyword Parameter Form Notes
:CURRent
:RANGe <dnpd>, <dnpd> Calibration range: <dnpd> = Low limit, High limit.
:RANGe? [<cpd> {LOW | HIGH}] Calibration range query.
:VOLTage <dpnd>, <dpnd> <dnpd> = low limit, high limit.
:VOLTage(?) [<cpd>{ LOW | HIGH }]
:UNIT? Response is VOLT (voltage) or CURR (current).
Query only.
:ACTual <dnpd>,<dnpd> <dnpd> = Amplitude, Angle. Note: Angle relative to
master phase.
:FREQuency? Query only of Actual frequency.
:TARGet <dnpd>[,<dnpd>,<dnpd>,<dnpd>,<dnpd>] <dnpd> = Target point.
Or,
<dnpd> =
Point,
Fund Freq,
Harmonic,
Amplitude,
Angle.
Note1: Angle relative to master phase.
Note2: Second form is only required when
changing target point.
:TRIGger?
:STORe
:DUMP? Dump all stores for active range:
Point,(<target data>,<actual data>),
<target data> =
Fund, Harm, Ampl, Angle
<actual data> =
Freq, Ampl, Angle
OUTPut
[:STATe](?) <bool> {OFF | ON | 0 | 1}
:ROSCillator
[:STATe](?) <bool> {OFF | ON | 0 | 1}
:SENSe(?) <bool> {OFF | ON | 0 | 1}
:DEFer(?)
[:STATe](?) <bool> {OFF | ON | 0 | 1}
:ACTion [<cpd> {APPLy | UNDO}]
:RAMP(?) {FAST | SLOW}
:TIME(?) <dnpd>
:RCLock(?) <dnpd>
:VOLTage
:NLIMit(?) <cpd> {LOW | HIGH}