Introduction
Electrical Specifications 1
1-13
Range
(Amps)
Frequency
Current
(Amps)
[4]
6105A & 6106A
1-Year Accuracy,
tcal
[3]
±5 °C ±(ppm of
output + μA)
6100B & 6101B
1-Year Accuracy,
tcal
[3]
±5 °C ±(ppm of
output + μA)
Open Loop
24-Hour Stability
±(ppm of output
+ μA)
[1][2]
4 - 10 A 49 100 191 240 75 50
45 Hz - 65 Hz
10 A 49 60 164 240 75 50
1 A - 4 A 65 200 191 240 75 100
10 A
16 Hz - 850 Hz
4 A - 10 A 65 200 191 240 75 100
8 A - 21 A 49 200 213 720 75 300
45 Hz - 65 Hz
21 A 49 120 189 720 75 300
2 A - 8 A 69 400 213 720 75 300
21 A
16 Hz - 850 Hz
8 A - 21 A 69 400 213 720 75 300
20 A - 50 A 49 500 213 1800 500 750
45 Hz - 65 Hz
50 A 49 300 189 1800 500 750
3.2 A - 20 A 74 1000 213 1800 500 750
50 A
40 Hz - 850 Hz
20 A - 50 A 74 1000 213 1800 500 750
8 A - 32 A 106 2800 265 2800 1000 1200
40 Hz - 450 Hz
32 A - 80 A 106 2800 250 2800 1000 1200
8 A - 32 A 112 2800 300 2800 1000 1200
80 A
450 Hz - 850 Hz
32 A - 80 A 118 2800 280 2800 1000 1200
[1] For ±1 °C and constant load and connection conditions.
[2] When Flicker, Fluctuating harmonics, Dip/Swell or Interharmonics are applied, “Open loop” stability specification must be added to
the 1-year accuracy specification as described in “Open and Closed Loop Operation”.
[3] tcal = temperature of last calibration.
[4] Output levels less than the range minimum can be set but are not specified.
[5] Settling time (TS) of 21 A, 50 A, and 80 A ranges depends on rms output as a proportion of full range and can be calculated from:
TS = %FR2 x 180 seconds. Settling magnitude will not exceed 50 ppm at FR.