6100B/6105A
Users Manual
5-50
Bit
Enable
Condition
8 Warm-Up Active.
9 Test Active.
10 Gate Trigger
Pending.
11 Energy Active.
STATus:OPERational:CONDition?
This query only command returns the contents of the Operation Condition register, which is
not cleared by the command. N. B. This register contains transient states, in that its bits are
not “sticky”, but are set and reset by the referred operations. The response to the query
therefore represents an instantaneous “Snapshot” of the register state, at the time that the
query was accepted.
This register is only used if the Energy Option is fitted.
Bit Condition Comment
8 Warm-Up Active Set 1 during the warm-up period.
Set back to 0 when the period ends.
9 Test Active Set 1 during the test period.
Set back to 0 when the period ends.
10 Gate Trigger
Pending
Set to 1 when waiting for the trigger.
Set back to 0 when trigger is received.
11 Energy Active Set 1 at the start of the Energy test.
Set back to 0 at the end of the test.
STATus:QUEStionable [:EVENt]?
Note
The 6100B does not set any bits in this register.
This command returns the contents of the Questionable Event register, clearing the register.
STATus:QUEStionable:ENABle(?) <dnpd>
This command sets the mask which enables those Questionable Event register bits which
are required to be summarized at bit 3 of the IEEE 488.2 Status Byte register.
STATus:QUEStionable:CONDition?
Note
The 6100B does not set any bits in this register.
This query only command returns the contents of the Questionable Condition register which
is not cleared by the command. N. B. This register contains transient states, in that its bits
are not “sticky”, but are set and reset by the referred conditions. The response to the query
therefore represents an instantaneous “Snapshot” of the register state, at the time that the
query was accepted.
STATus:PRESet