Section 5: Performance Verification and Safety Testing
2025653-048 Revision B Responder
2000 Page 73
Leads Name Limit
8 Patient Leakage Current AC (OS) 500 µA
8 Patient Leakage Current AC (NC) 100 µA
8 Patient Leakage Current AC (OE) 500 µA
8 Patient Leakage Current AC OSRM 500 µA
8 Patient Leakage Current AC NCRM 100 µA
8 Patient Leakage Current AC OERM 500 µA
1-2 Patient Auxiliary Current AC (OS) 50 µA
1-2 Patient Auxiliary Current AC NC 10 µA
1-2 Patient Auxiliary Current AC OE 50 µA
1-2 Patient Auxiliary Current AC OSRM 50 µA
1-2 Patient Auxiliary Current AC NCRM 10 µA
1-2 Patient Auxiliary Current AC OERM 50 µA
3-7 Patient Auxiliary Current AC (OS) 50 µA
3-7 Patient Auxiliary Current AC (NC) 10 µA
3-7 Patient Auxiliary Current AC (OE) 50 µA
3-7 Patient Auxiliary Current AC OSRM 50 µA
3-7 Patient Auxiliary Current AC NCRM 10 µA
3-7 Patient Auxiliary Current AC OERM 50 µA
1-2 Mains on Applied Part (SFC) 50 µA
1-2 Mains on Applied Part (SFCRM) 50 µA
3-7 Mains on Applied Part SFC 50 µA
3-7 Mains on Applied Part SFCRM 50 µA
8 Mains on Applied Part SFC 5000 µA
8 Mains on Applied Part SFCRM 5000 µA
1-2 Patient Leakage Current DC (OS) 50 µA
1-2 Patient Leakage Current DC (NC) 10 µA
1-2 Patient Leakage Current DC (OE) 50 µA
1-2 Patient Auxiliary Current DC OSRM 50 µA
1-2 Patient Auxiliary Current DC NCRM 10 µA
1-2 Patient Auxiliary Current DC OERM 50 µA