Section 5: Performance Verification and Safety Testing
2025653-048 Revision B Responder
2000 Page 79
Enclosure (Chassis) leakage
Perform the following tests, with the applicable switch settings on the “Safety Tester” and record the results in the
applicable test form(s).
UUT power switch in the “ON” position.
TP1 connected to accessible metal parts (see tables below for individual test points).
S1: DPDT
Polarity
a) Normal
b) Reverse
S2: SPDT
Neutral
a) Closed
b) Open
S3: SPDT
Ground
a) Closed
b) Open
S4: SPDT
Test point
a) Internal
b) TP1
Position Position Position Position
Maximum
Limits
(uA)
Test Description
a b a b a b a b
EN/IE
C
UL
Enclosure (Chassis) leakage current measurements:
TP1 connected to “RS-232 DB9 shell and Metal part of Printer.
Polarity: Normal
Neutral: Normal (Closed)
Ground: Normal (Closed)
X X X X
100
100
Polarity: Normal
Neutral: Open
Ground: Normal (Closed)
X X X X
500
300
Polarity: Normal
Neutral: Normal (Closed)
Ground: Open
X X X X
500
300
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Open
X X X X
500
300
Polarity: Reverse
Neutral: Open
Ground: Normal (Closed)
X X X X
500
300
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Normal (Closed)
X X X X
100
100