DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
3-102 Section 3-10 - Connectivity Setup
3-10-3-8 Configuring the Vivid i n/ Vivid q N and Network PC for a Peer-to-Peer Connection
NOTE: The following steps are performed on the Vivid i n/ Vivid q N scanner.
The instruction Select = trackball to the specific item and press the Set button on the control panel.
1.) On the Vivid i n/ Vivid q N control panel, press the Config button.
2.) From the System Configuration window, click the Connectivity tab at the bottom of the window.
The Connectivity dialog box opens.
3.) Select the TCPIP Tab.
4.) Trackball to the Advanced tab and press the Set button on the control panel.
5.) Select the Local Area Connection, as shown in Figure 3-105.
6.) Press the Menu button on the control panel.
7.) From the drop-down menu, select Properties.
The Local Area Connection Properties dialog box opens.
Figure 3-105 Connectivity Dialog Box - TCPIP Advanced Tab
Figure 3-106 Local Area Connection Properties