DIRECTION FQ091013, REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL
Chapter 3 - System Setup 3-115
c.) Edit the name and/or the IP address of the server, see Figure 3-123.
d.) Click OK to save the new settings and close the Edit Name dialog box.
e.) Click OK to close the IPs dialog to return to the Properties dialog box.
f.) Enter the DICOM server AE Title. This entry is case-sensitive and must match exactly.
g.) Enter the DICOM server port.
h.) For some DICOM Servers, the default Time-out setting (30) is too low. Change if necessary
by typing a new value (seconds).
i.) Click OK to close the Worklist Properties dialog box and save the changes.
Figure 3-123 Edit Name and/or IP address
Figure 3-124 DICOM Server’s Properties Dialog Box
Name
IP address
AE Title
Port No