Heska Element DC - Work List Selection Setting; Coefficients;Test Settings <Admini.>

Heska Element DC
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. Touch YES or NO in the [Keep sequence No.:].
5. Touch REGISTER.
The sample No./ID settings are registered, and then the
display returns to the DISPLAY SETTINGS menu dialogue.
NOTE: To cancel setting, touch CANCEL.
6.9.6 Work List Selection Setting
This mode is used to set the work list index display.
1. Touch DISPLAY SETTINGS in the ADMINISTRATOR MODE FUNCTION menu dialogue.
The display settings menu dialogue is displayed.
NOTE: For information about how to display the ADMINISTRATOR MODE FUNCTION menu dialogue, refer to Section 6.1.2.
2. Touch WORK LIST SELECTION SETTING in the menu
dialogue.
The work list selection setting dialogue is displayed.
3. Touch SAMPLE ID/PATIENT NAME or PATIENT ID/PATIENT
name.
4. Touch REGISTER.
The work list selection setting is registered, and then the
display returns to the DISPLAY SETTINGS menu dialogue.
NOTE: To cancel setting, touch CANCEL.
6.10 Coefficients/Test Settings <Admini.>
This mode is used to set the following.
Correlation coefficients
Lot compensation coefficients
Dilution factor
Spotting count
Print sheets of test results
103