3.7.4 List for Daily Maintenance after Use
According to Section 3.7.5, please perform the daily maintenance after use.
Item Contents Reference Section
Cleaning disposal box Empty the disposal box of used slides and tips. 3.7.5 (2)
Removing used consumables
Dispose of used mixing cup and remaining reference fluid and remove
remaining sample.
3.7.5 (3)
Cleaning spotting part and sample rack Clean the spotting part and around the sample rack. 3.7.5 (4)
3.7.5 Turning the Power Off (Daily Maintenance after Use)
1. Press Power button or Power OFF on the LCD touch panel.
WARNING
Do not turn off the power during measurement process and
mode operations. Otherwise, physical damage may occur.
If the power is turned off during measurement process,
load consumables (tips, etc.) and perform the tests from the
beginning. Be sure to replace the mixing cup for tests that
require dilution.
NOTE: If analyzer will not be used for a long time, depress the
[
] side of the power switch.
2. Remove the disposal box and dispose of the tips and slides.
3. Dispose of used mixing cup and remaining reference fluid
and remove remaining sample.
Element DC
MIX CUP
d
c
RE
b
D
a
16x100
Discard
Remove
29