4. Clean around the sample rack and the spotting part.
Use swabs moistened with ethyl alcohol as a disinfection for
cleaning.
WARNING
When disinfecting (cleaning) the analyzer, always follow
biohazard procedures (e.g., wearing gloves, lab coat, and safety
goggles).
If any part of the body comes in contact with contaminated
parts, immediately rinse the contaminated body part
thoroughly under running water and then use ethyl alcohol as
a disinfectant. Seek medical assistance if necessary.
WARNING
As contaminated swabs used for cleaning the analyzer are
infectious waste, process the waste correctly in compliance
with any applicable regulations in your country, such as by
incineration, melting, sterilization or disinfection.
5. If the LCD touch panel is dirty, use a soft cloth or a cloth damped with a small amount of ethyl alcohol to clean by
wiping gently.
6. Close the sampler cover.
3.7.6 Sample Racks
Sample racks depend on sample tubes to be used.
IMPORTANT
Aassembling incorrectly may cause adverse effects on test results due to incorrect suction or analyzer damage;
assemble the sample racks correctly.
1. Sample rack assembly for a 1.5 mL sample tube.
1.5mI
MIX CUP
a
b
Place label “1.5 ml”.
Insert the tube holder “3” (for 1.5 mL tubes)
in the hole.
Insert a pin “7” in the left
(back view) hole.
30