5.2.6 Transfer Errors
Error Code Error Message Error Status Solution
E0200
E0205
E0208
E0209
Feed motor ERR The disposal box may be full of slides,
or slides may remain in the incubator
or the spotting part.
Turn the power switch off, and perform
the following:
1. Empty the disposal box.
2. Incubator cleaning (Refer to
Section 4.3).
3. Spotting part cleaning (Refer to
Section 4.3).
If the error still occurs, please contact
Heska's Technical Support Services for
assistance.
E0206
E0207
Feed motor ERR
Rerun the tests.
E0210 Feed motor ERR
E0600
E0601
Incubator motor ERR
E0602 Incubator motor ERR
E0603 Incubator motor ERR
Check Disposal box.
E0700
E0701
Probe motor ERR
E0702 Probe motor operation ERR
E1600
E1601
E1603
Transfer change motor ERR
E1602 Transfer change motor ERR
E1700 Eject motor ERR
The sensor failed to turn ON.
E1701 Eject motor ERR
The sensor failed to turn OFF.
E1702 Eject motor ERR
E1703 Eject motor ERR
Check Disposal box.
5.2.7 Temperature Control Errors
Error Code Error Message Error Status Solution
E1500 Temperature control ERR
Check room temperature.
Clean air filter.
IMPORTANT
Measured values with “+“ or “–“ marks
may NOT be accurate. Perform the
following troubleshooting and rerun
the tests.
When “+” or “–” marks appear on test
results or the following errors occur:
E1500, E1501, E1512
(CM incubator temperature)
E1504, E1510, E1513
(ISE incubator temperature)
E1521 (Environmental temperature)
Incubator temperature is not within
the controlled range.
Perform the following troubleshooting.
a. Make sure the room temperature is
between 59–89°F (15–32°C). If it is out
of the range, air-condition or heat the
room.
b. Turn the power switch off.
c. Clean the air filter. (Refer to Section
4.2)
d. After the incubator has cooled down
(about 10 minutes), turn the power
switch on.
If the error still occurs, please contact
Heska's Technical Support Services for
assistance.
E1501
E1512
Temperature control ERR
Check room temperature.
Turn power SW OFF.
E1504 ISE temperature control ERR
Check room temperature.
Clean air filter.
E1510
E1513
ISE temperature control ERR
Check room temperature.
Turn power SW OFF.
E1521 Room temperature WARNING
Check room temperature.
Clean air filter.
69