1. Touch COEFFICIENTS/TEST SETTINGS in the
ADMINISTRATOR MODE FUNCTION menu dialogue.
The coefficients/test settings menu dialogue is displayed.
NOTE: For information about how to display the
ADMINISTRATOR MODE FUNCTION menu dialogue, refer to
Section 6.1.2.
2. Touch CORRELATION COEFFICIENTS in the menu dialogue.
The correlation coefficients TOP dialogue is displayed.
By touching or , you can see all correlation coefficients
presently input.
NOTE: Test name, sample type and correlation coefficients (a, b)
are displayed.
NOTE: Test names are displayed in ascending order of test code.
Sample types are displayed in order of: [P/S, W].
NOTE: By touching
, all correlation coefficients presently
input are printed out.
NOTE: To cancel setting, touch BACK.
NOTE: When touching RESET ALL TESTS, the reset confirmation
dialogue appears.
Touch OK to reset all correlation coefficients. Touch CANCEL to
cancel a reset.
The display returns to the correlation coefficients TOP dialogue.
3. Touch DATA INPUT.
The test names selection dialogue is displayed.
By touching or , you can see all test names.
NOTE: All test names presently recorded in the analyzer memory
are displayed.
6.10.1 Correlation Coefficients
This mode is used to input, reset or print out correlation coefficients (a, b) for each test and for each sample type (P/S, and
W). For further details, refer to the Description of the Correlation Function at the end of this section.
IMPORTANT
Incorrect inputs for (a, b) will cause incorrect test results. Make sure that the inputs (a, b) are correct using this mode.
104