Printout Message
Description and status Solution
E
Impossible to calculate the concentration of the sample because of some
interruption (e.g., slide jam) of testing.
The test results will be printed out as “****”.
Check other errors previously
occurred, and perform the
necessary troubleshooting.
& Abnormal high testing value.
Rerun the tests.
Perform tests that require
dilution, following the directions
on the Instructions for Use of the
slides.
#
The valid term of the slide has expired.
IMPORTANT
The testing value may NOT be accurate. Refer to the related troubleshooting
pages and rerun the test.
IMPORTANT
Expired slides cannot be used. Use valid slides for expiration date.
-
¥ Unspotted slide. Rerun the tests.
C Testing value is a result in control mode.
The control mode can be
canceled by the control mode in
MODE FUNCTION.
ERR = xxx
Errors related to ISE test results.
("x" indicates the numeric number from 0 to 8.)
IMPORTANT
The testing value may NOT be accurate. Refer to the related troubleshooting pages
and rerun the test.
Refer to Section 5.2.11
(Supplement) About Measurement Range
The measurement range (determination range) and the reference interval differ in value for each test. The relation
between the measurement range and the reference interval is shown below. For details, refer to the Instructions for Use for
the slides.
NOTE: The values of the reference intervals can be set by using the [Reference Interval Settings] of MODE FUNCTION (Refer
to Section 6.9.3).
64