2. W141
Error Code Error Message Error Status Solution
W141 Unacceptable ISE test
ISE tests cannot run.
Turn the power SW OFF and ON.
Rerun the tests from the beginning.
E0530—E0532 occurred; analyzer has not
been turned off (reset).
Turn the power switch off and on, and
then rerun the tests from the beginning.
3. When an error code (ERR = xxx) is printed out along with the test results:
During ISE tests, the following will be checked.
Check Menu Description
Drift check Checks abnormal time course (voltage) during testing
Impedance check Checks the slide impedance after the test completed
Over range check Checks that the test results are within the measurement range
If any errors are detected, [ ERR = xxx ] will be printed out along with the test results. The 3-digit figures are error codes
for Na (sodium), K (potassium), and Cl (chloride).
NOTE: “ x ” indicates a numeric number from 0 to 8.
Error code description
Code Description
0 No problem
1 Drift error
2 Impedance error
3 Outside of the measurement range (This is not an analyzer malfunction.)
4 Drift error and impedance error
5 Impedance error and outside of the measurement range
6 Drift error and outside of the measurement range
7 Drift error, impedance error, and outside of the measurement range
8 Impossible measurement
4. W142
Error Code Error Message Error Status Solution
W142 Unacceptable ISE test
ISE tests cannot run.
ISE probe unit not loaded inside
analyzer for attempted ISE slide
test.
—
ERR = x x x
Error code for Cl
Error code for K
Error code for Na
72