22
ml
5
0
.
φ
13
7
5
×
0.5ml
13
75
×
Blood collecting tube
or FUJI Sample Tube
A
B
E
C
D
Sample rack
Slide cartridge
Blood collecting tube
or FUJI Sample Tube
A
B
E
C
D
Sample rack
Slide cartridge
STOP
0
3
B D
C
E
MYID001
..0160503010
..0160503009
1
8
2
7
3
4
5
6
e. Make sure the sample rack and sample tube size match.
NOTE: To avoid a suction error or malfunction, use a sample tube
that matches the label on the sample rack.
NOTE: Refer to Section 3.8 for sample tubes that can be used.
f. Remove the cap of the sample tube that contains sample
and then place it into the specific sample rack.
NOTE: Make sure there are no bubbles in the sample in the sample
tube. Use a centrifuge to remove bubbles.
IMPORTANT
Be sure the sample and its measurement slides are in the correct
positions.
4. Registering sample information.
Touch a sample rack position to display the sample
information registration screen.