EasyManua.ls Logo

Hettich HettCube 200 - Page 99

Hettich HettCube 200
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
99 / 174
en
12.2020 / Rev. 02
AB66005
Operation
Class 3.1
Fig. 152: ‘Temperature guard’ menu,
class 3.1
Triggering example, class 3.1
Class 3.2
switched off; for class 3.2, the cooling. As soon as the actual tem-
peratur
e is back within the limits, the incubator will continue to regu-
late. The temperature guards serve as device protection and sample
protection. The temperature limits can limit the standard temperature
range.
In the "Temperature guard" settings, temperatures for classes 3.1
and 3.2 can be input.
For 3.1, up to +70 °C
For 3.2, from -10 °C up to +70 °C
If classes 3.1 and 3.2 are set, they must lie at least 4°C / 8°F apart.
It is only sensible to use a temperature guard of class
3.1 in heating mode with monitoring temperatures
which lie above room temperature.
The temperature specication value can be changed by tapping it.
The setting ex works is +70 °C. This temperature lies outside of the
temperature range of the incubator. Class 3.1 here acts like device
protection, but is not indicated as being active on the main screen. If
the setting is changed, the class 3.1 is indicated as being active on
the main screen.
Target temperature +37 °C, class 3.1 at +40 °C
Fig. 153: Diagram of ‘Temperature guard’, class 3.1
A Set temperature
B Temperature specication, class 3.1
C Deviation from the target temperature starting from time C
D Class 3.1 active starting from time D
It is only sensible to use a temperature guard of class
3.2 in cooling mode with monitoring temperatures
which lie below room temperature.

Table of Contents

Related product manuals