Brevera Breast Biopsy System User Guide
Chapter 7: Biopsy
Page 72 MAN-04303-002 Revision 006
7. Get a saline bag and remove the protective cap. (A 250 cc saline bag is
recommended.)
8. Remove the saline tubing from the biopsy needle packaging.
9. Insert the spike into the saline bag.
WARNING!
Make sure you use an aseptic technique when spiking the saline bag to
prevent contamination.
10. Place the saline bag on the saline bag hook on the left side of the console.
11. Route the saline tubing from the spike through the saline tubing notch,
counterclockwise around the saline tubing management plate, and through the saline
tubing conduit.
12. Insert the larger diameter section of the saline tubing into the saline pinch valve.
Make sure the saline tubing is fully seated.
Figure 45: Placement of the Saline Tubing in the Saline Pinch Valve
13. Insert the vacuum tubing into the guide at the top of the saline tubing management
plate.