EasyManua.ls Logo

HP 4062UX - Page 142

HP 4062UX
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
b. Figure 6-7 shows an example of an HP 16346C calibration data entry page. You can
move the eld p ointer by using the arrow keys or the
4
Tab
5
key. The data at the eld
pointer is displayed in the display area. If you wantto change the displayed data,
type in new data, then press
4
Return
5
. After entering the calibration data, select the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SAVE/EXIT
softkey to return to the calibration mo dule menu page.
Use the same metho d if you need to enter calibration data for other calibration
mo dules.
d a
c b
* HP 4062F PERFORMANCE VERIFICATION *
| | | DC | 1 KHz | 10 KHz | 100 KHz | 1 MHz |
| 100 ohm | R | 100.00 ohm | | | | |
| |C| | | | | 0.0000 pF |
| 1 Kohm | R | 1.0000 Kohm | | | | |
| |C| | | | | 0.0000 pF |
| 100 pF | C | | | 100.00 nF | 100.00 nF | 100.00 nF |
| 1000 pF | C | | 1000.0 pF | 1000.0 pF | 1000.0 pF | 1000.0 pF |
|10nF |C| | 10.000 nF | 10.000 nF | 10.000 nF | |
| 100 nF | C | | 100.00 nF | 100.00 nF | | |
Enter a value ( 100 ohm ).
100.00
User 1 Caps Running
aaaaaaaaaaaaaaa aaaaaaaaaaaaaaa aaaaaaaaaaaaaaa aaaaaaaaaaaaaaa
CANCEL
aaaaaaaaaaaaaaa aaaaaaaaaaaaaaa aaaaaaaaaaaaaaa aaaaaaaaaaaaaaa
SAVE /
EXIT
Figure 6-31. Example of Calibration Data Entry P
age: HP 16346C
Note
Calibration data is stored in the /usr/p cs/diag/CAL PV le and is used each
time the PV software is run.
c. After entering the necessary calibration data, select the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Start-up Menu
softkey on
the calibration mo dule menu page.
d. Select softkey of your system calibration mo dule.
6-30 System Verification

Table of Contents