Making Automatic Measurements
Freq Domain Menu
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
Pk threshold
The Peak Threshold function sets the threshold level for peak searches.In
order to be considered a peak, a local max in the FFT spectrum must be
above the threshold level and must go up and down at least one-eighth of full
scale (one division).
Key Path
4
Shift
54
More Meas
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Freq Domain
NNNNNNNNNNNNNNNNNNNNNNNNNN
FFT Freq
(or
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FFT Magnitude
or
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FFT
1Freq
or
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FFT
1Magnitude
)
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pk
threshold