Eye, Mask and Eyeline Mode Measurement Tutorials
Making Eye Diagram Measurements
Positioning the waveform
To make Eye Diagram measurements, the data waveform must be properly
positioned on the instrument display. The measurement algorithm needs at
least two full crossing points on screen to be able to make the measurements.
You should also have about four divisions of vertical scale. Refer to
Figure 5-2.
Figure 5-2. Position of data waveform.
T
o
accomplish
this
you
need
to
set
the trigger
,
vertical
scale
,
and
the
time
base
.
5-5