General Purpose Oscilloscope Menus
FFT Menu
Amplitude measurements
For best amplitude accuracy on peaks
1. In the Channel menu, make sure the source impedance and probe attenuation are set correctly for
your application.
2.
In
the
Channel
menu,
set
the
Scaling
for
almost
full
scale
deection.
3.
In
the FFT
menu, use
the attop
window.
4.
Press
4
Shift
5
4
Mo
re
meas
5
on
the
ke
ypad, then
make an
FFT
magnitude
measurement.