CONNECTORS
frequency to
+lO
V at the specified maximum frequency, with a
proportional voltage for frequencies between the specified minimum
and maximum. When the swept signal generator is in manual sweep
operation the sweep output voltage is a percentage of the span.
Minimum load impedance is 3 kR.
STOP SWEEP IN/OUT stops a sweep when this input is pulled
low. Retrace does not occur, and the sweep resumes when this
input is pulled high. The open circuit voltage is TTL high and is
internally pulled low when the swept signal generator stops its sweep.
Externally forcing this input high will not cause damage or disrupt
normal operation.
10 MHz REF INPUT accepts a 10 MHz
flO0
Hz, 0 to
+lO
dBm
reference signal for operation referenced to an external time base.
Nominal input impedance is 50
CR.
10 MHz REF OUTPUT provides a 0
dBm,
10 MHz signal derived
from the internal frequency standard of the swept signal generator.
This input is a 50 fl connector that can be used as the master clock
reference output for a network of instruments.
TRIGGER INPUT activated on a TTL rising edge. Used to
externally initiate an analog sweep or to advance to the next point of
a step list or a frequency list.
TRIGGER OUTPUT Produces a 1
pus
wide TTL-level pulse at 1601
points evenly spaced across an analog sweep, or at each point in a
step list or a frequency list.
VOLTS/GHz
supplies a voltage that is proportional to the RF
output frequency, with a ratio of 0.5 volt output for every 1 GHz
of RF frequency (factory setting). This ratio is switchable to
either 0.25 or 1 volt. The switch is located on the Al2 SYTM
assembly, see Adjustments in the HP 8360 B-Series Swept Signal
Generator/HP 8360 L-Series Swept CW Generator Service Guide for
information. This output is designed to drive into 2 kR or greater.
Z-AXIS
BLANK/MKRS
supplies a positive rectangular pulse
(approximately
+5
V into 2
k0)
during the retrace and switch
points when the swept signal generator is sweeping. This output
also supplies a -5 V pulse when the RF output is coincident with a
marker frequency.
Multi-pin Connectors
AUXILIARY INTERFACE connector provides control signals to the
HP 8516A S-parameter test set switch doubler. This connector is a
25-pin D-subminiature receptacle located on the rear panel. It is also
used for dual swept signal generator measurement systems (two-tone
systems), refer to Step Control
Master
for more information.
C-6 Operating and Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8360 B Series and is the answer not in the manual?

HP 8360 B Series Specifications

General IconGeneral
BrandHP
Model8360 B Series
CategoryPortable Generator
LanguageEnglish

Related product manuals