If this is ON, each time a frequency or power is changed,
CALibration:AM[:EXECute]
is attempted.
l
CALibration:AM[:EXECute]
When AM is on and the swept signal generator is in the CW or
manual mode, the swept signal generator performs a modulator
calibration as long as power sweep is not active.
l CALibration:PEAKing:AUTO ONlOFFlllO
l CALibration:PEAKing:AUTO?
Sets and queries the automatic peaking function. If AUTO is ON, then
a peak is done at regular intervals automatically. After
*RST,
the
setting is OFF.
l
CALibration:PEAKing[:EXECutel
Peaks the SYTM.
l CALibration:PMETer:DETector:INITiate?
IDETectorlDIODe
Initiates the specified calibration. These calibrations require the use
of an external power measurement. Once initiated, the swept signal
generator sets up for the first point to be measured, and responds to
the query with the frequency at which the power is to be measured.
The parameters mean:
IDETector
DIODe
Initiates a calibration of the internal detector logger
breakpoints and offsets.
Initiates a calibration of an external detector’s logger
breakpoints and offsets.
l CALibration:PMETer:DETector:NEXT?
<num>[lvlsuffix]
The parameter is the measured power that is currently produced by
the swept signal generator. You must supply this parameter after
measuring the power using an external power meter. The query
response is issued after the swept signal generator processes the
supplied parameter and settles on the next point to be measured.
The query response is:
>o
The frequency [in Hz] that is currently produced.
0
The calibration is complete.
<O
An error has occurred and the calibration is aborted.
l CALibration:PMETer:FLATness:INITiate?
USERlDIODelPMETerlMMHead
Initiates the specified calibration. These calibrations require the use
of an external power measurement.
Once initiated, the swept signal
generator sets up for the first point to be measured, and responds to
the query with the frequency at which the power is to be measured.
The parameters mean :
USER
Initiates a calibration at all of the user flatness points.
S-28 Operating and Programming Reference