Multitron - Incubation Shaker
Operation
25 February 2019 Page 93 of 146
Checking the timer for the second phase
7. Tap the TIMER key to get to the time settings for the second
phase (Time 2).
On the lower alphanumeric display appears the time entered
last or 00:00 if no time has been entered.
In the separate timer display area the signs Timer and Phase
2 are lit up.
8. Make sure the second timer is turned off. If not, use the
ON/OFF key in the lower setting key area to turn the timer
(Time 2) off.
Entering the setpoints for the second phase and switching on
the parameters
After making sure, that the second timer (Time 2) is switched off,
the different parameters for phase 2 can be selected using the Se-
lect key and afterwards be switched on.
INFORMATION
If the timer should be used to stop the cultivation process after a
defined time, all parameters for phase 2 can be switched off.
This causes the parameters to switch off after the set time for
phase 1 has elapsed and therefore to stop the cultivation pro-
cess.
9. Tap the Select key to select the desired parameters.
In both display areas the first integral pair of parameters
(Temp and RPM) appears with the last setpoint values en-
tered or the message OFF.
The sign Setpoint indicates, that the setpoints can be entered.