VIsualEyes 515/525 - Additional Information Page 86
Figure 8.2.4 Saccade subtest settings
Test Name: Name of subtest
Duration: Length of time to perform the test based on the number of saccades
Fixation light: Determines when the fixation light in the goggles is displayed. Manual will leave the operator
to determine when the fixation light should be displayed. Timed will have the fixation light come on
automatically at the designated time. Disabled (default) will block the fixation light from turning on in the test.
Save Eye Recording: Choice to save eye video recordings.
Number of Jumps: Defines how many saccade jumps the patient will follow.
Maximum target movement: Define maximum degrees target can move relative to previous target position.
Minimum target movement: Define minimum degrees target can move relative to previous target position.
Compensate for drift: If enabled, the software will adjust the patient’s response to better align with the
target.
Extend end of Recording: If enabled the test will return the target back to center for two seconds. Not
recommended if the tests are run automatically.