Note: You cannot set the time more than 24hours in the past.
Update from cloud
1.
Restart your EVOS
™
M5000 instrument before an update to optimize the update process.
2.
To update the instrument software and
firmware from the Cloud, sign in to your Connect
account
(“Connect to the Thermo Fisher
™
Connect Platform” on page23).
3.
On the
Service tab, click Update from Cloud.
The instrument automatically searches your Connect
account, for the update.
4.
If a new update is available and the appropriate files are detected, the software pauses for a few
moments and then r
estarts.
5.
When a Windows
™
OS Setup dialog box opens, click Next
to proceed with the software update.
6.
The screen displays the update progress. When the update is complete, the system restarts and
the M5000 softwar
e automatically boots up again.
Update from USB
1.
Download the latest software from thermofisher.com/evos to a fresh USB flash drive with at least
200MB of available space.
2.
Restart your EVOS
™
M5000 instrument before an update to optimize the update process.
3.
Plug the USB
flash drive containing the newest software update into the EVOS
™
M5000instrument.
4.
On the
Service tab, click Update from USB.
5.
Locate the update
file from your USB Drive, then click Open.
6.
When a Windows
™
OS Setup dialog box opens, click
Next to proceed with the software update.
7.
The screen displays the update progress. When the update is complete, the system restarts and
the M5000 softwar
e automatically boots up again.
Chapter11Configure instrument settings
Service
11
106
EVOS
™
M5000 Imaging System User Guide