6.
Attach the tool to the new light cube and lower the cube into position so that the electronic
connection aligns properly (facing the back of the microscope) and the cube sits squarely in place
with the label facing toward the front.
7.
Unscrew the light cube tool from the cube, then use it to gently tighten the two slotted screws flush
with the ridges on the light cube.
Change the objectives
To customize your EVOS
™
M5000 Imaging System, you can add and remove objectives to fit the
instrument’
s functionality to your specific research needs.
Procedure for objective change
1.
Remove the objective you want to replace from the objective turret. You may need to move the
stage so that the objectives are accessible. Note the indicated position (1–5) of the removed
objective on the turret (red arrow).
Chapter12Instrument care and maintenance
Change the objectives
12
110
EVOS
™
M5000 Imaging System User Guide