Note: Make sure to set aside packaging and foam for future transport and storage. Re-install the stage
lock pin and the light cube shipping restraint before moving or transporting the instrument. Always
ensure that the instrument is properly cushioned and braced to prevent damage.
Remove shipping restraints
The EVOS
M5000 Imaging System is equipped with two shipping restraints (stage lock pin and light
cube shipping restraint) to protect the instrument from shock and vibration during transport. You must
remove the shipping restraints before you power on the EVOS
M5000 Imaging System.
2
3
1
Figure1Before removing the stage lock pin.
1
Stage lock pin
2
Light cube shipping restraint
3
Light cube tool
1.
Pull firmly to remove the stage lock pin.
2.
Using the Y-axis stage positioning knob, move the stage back to obtain access to the light cube
shipping r
estraint, which is centered under the back of the stage.
Note:
The light cube shipping restraint is secured with the light cube tool to the blank light cube
installed in the light cube turr
et. Used together, they immobilize the light cube turret to protect it
during transport.
Chapter3Installation