EasyManua.ls Logo

LAR Process Analysers QuickTOC purity - Calibration Results

LAR Process Analysers QuickTOC purity
196 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
98 User Manual QuickTOC
®
purity
ATEX 2 04E3920
7 How to Work With the Analyser
7.2 Configuration LAR | PROCESS ANALYSERS AG
Fig. 59: Calibrate Screen - Show Details (Example)
Fig. 60: Calibrate Screen - Calibration Graph (Example)
7.2.4.2 Calibration Results
After a successful calibration (Chapter 7.2.4.1 from page 95), the results are displayed on the "Calibra-
tion Results" screen (Fig. 61, page 99). To the right, the values of the previous calibration are displayed.
Please activate the calibration via the button "Activate the calibration results". The result is saved. Fur-
thermore, this screen displays the "carrier current setpoint".
Back to Start of Calibration
Back to Start of Calibration

Table of Contents