04E3920 User Manual QuickTOC
®
purity
ATEX 2 139
8 Care and Maintenance
8.3 Care and Maintenance Tasks
LAR | PROCESS ANALYSERS AG
8.3.2 Actions
Use a copy of the "Care Protocol (Analyser)" for the care actions, which you will find in Kapitel 13.3 auf
Seite 178.
1 year Replace reactor tube or
change reactor filling (if
necessary)
Maintenance Chapter 8.3.17 on
page 147
Replace reactor seal (if
necessary)
Maintenance Chapter 8.3.18 from
page 149
Injection needle (if neces-
sary) and tube for
Replace injection needle
Maintenance Chapter 8.3.19 from
page 151
Replace loop tubing Maintenance Chapter 8.3.19 from
page 151
Change sample drain tubes
(if necessary)
Maintenance Chapter 8.3.20 from
page 152
Replace quartz wool filling Maintenance Chapter 8.3.21 from
page 152
Replace pre-detector gas
filter 0,1 μm
Maintenance Chapter 8.3.22 from
page 153
If neces-
sary
Check the Viton hoses,
clean or replace if neces-
sary
Care Chapter 8.3.23 from
page 153
Interval Measure
Task
completed
Notes
1 week Check pump tubes and
clean if necessary
Check inlet and outlet of the
control unit for impurities
Check ventilator for impuri-
ties
Interval Task Task Type Chapter