180 User Manual QuickTOC
®
purity
ATEX 2 04E3920
13 Logs and Protocols
13.5 Functional Test Protocol (Analyser) LAR | PROCESS ANALYSERS AG
13.5 Functional Test Protocol (Analyser)
Visual inspection Criteria OK Measure
Analyser status • Zero signal between
0 - 0,1 FSR
• Carrier gas on / off:
approx. 30 l/h
Contact support
______________________
______________________
______________________
Tightness test • Carrier gas rate falls to
< 5 l/h
Contact support
______________________
______________________
Checking the measured
values
• Measured values
correspond to the
calibration standard
Contact support
______________________
______________________
Checking the measurement • Measurement takes place
without problems
Contact support
______________________
______________________
Date: Signature