8 Care and Maintenance
8.2 Protocol for Visual Inspection (Analyser)
LAR | PROCESS ANALYSERS AG
8.2 Protocol for Visual Inspection (Analyser)
Proceed as follows:
1. In user level 2, call up the "Status screen" display.
2. Check if the zero signal is between 0 - 0,1 FSR.
Visual Inspection Criteria OK Measures
Interval: 1 week
Zero signal • 0 - 0,1 FSR
Check soda lime
Contact support
Carrier gas • Carrier gas on / off:
ca.13,5 - 15,5 l/h
Check fittings
Contact support
Injection system • no air bubbles in the glass Rinsing
Glass components • no impurities Cleaning
Canister and supply tubes • levels > 1 liter
• no impurities
• normale elasticity
Top up canisters
Clean canisters
Replace tube
Input and drain tubes • no impurities
• normal elasticity
Replace tube
_____________________
Check inlet and outlet of
the control unit for impu-
rities
• no impurities
• no corrosion
Contact support
Check ventilator for
impurities
• no impurities
• no corrosion
Clean ventilator
Interval: 3 months
Injection tube
• no impurities
• normal elasticity
Cleaning
Replace tube
Tube cassette pump
and sample pump
• no humidity
• rollers run smoothly
• no impurities
• normal elasticity
Cleaning
Set tubes forth
Replace tube
Contact support
Sample and pump
tubes
• no impurities
• rollers run smoothly
Cleaning
Contact support
Viton tubes
• no impurities
• normal elasticity
Cleaning
Replace tube
Acid trap
• at least 1/3 of the zinc is
shiny
• at least 1/3 of the brass
wool is yellow
Replace filling
Replace acid trap
Quartz wool filter
• no humidity
• no discolorations
Replace filling
Replace filter
Date: Signature: