04E3920 User Manual QuickTOC
®
purity
ATEX 2 135
LAR | PROCESS ANALYSERS AG
8 Care and Maintenance
8 Care and Maintenance
Only minor effort is required to service and maintain the analyser. This section shows you the best way
to look after your analyser to guarantee trouble-free operation. The documentation of maintenance and
service work is a precondition for any warranty and guarantee claims, and also represents a valuable
aid in locating resolutions when malfunctions occur (Chapter 10 on page 163).
Care and Maintenance:
Care measures are application-dependent and are recommended by LAR. The effort for
maintenance work is about 30 minutes / week.
Maintenance includes replacement of consumables and consumables. The cost of maintenance is
about 5 - 10 min / week.
After maintenance and servicing, some functional tests must always be performed to check analyser
status
Notice
The scope of analyser maintenance and care work depends on the application.
All maintenance, care action and intervals, pertain to the most demanding of applica-
tions and are to be understood as recommendations from LAR.
Danger
Ex-Protection
Before servicing or repairing, turn on the bypass of the pressurized enclosure system
on the right side.
Stay away from the analyser when an explosive atmosphere prevails. Maintain and
repair the analyser only when it is ascertained that there is no explosive atmosphere
around the analyser.
Notice
If you have any questions about maintenance and / or service, contact LAR Technical
Support (Chapter 15.1 on page 267).
As part of customer support, LAR Technical Support offers customizable mainte-
nance contracts and device-specific training to extend your know-how. Further in-
formation at: www.lar.com.