178 User Manual QuickTOC
®
purity
ATEX 2 04E3920
13 Logs and Protocols
13.3 Care Protocol (Analyser) LAR | PROCESS ANALYSERS AG
13.3 Care Protocol (Analyser)
Interval Measure
Task
completed
Notes
1 week Clean vessel for calibration fluid
and replace calibration standard
Check pump tubes and clean if
necessary
Check inlet and outlet of the con-
trol unit for impurities
Check ventilator for impurities
3 months Check reactor pipe end and clean
if necessary
Inspection of the gas cooling
pipes
Check pump cassettes and clean
if necessary
Check the rollers of the pump
head for ease of movement
Checking the measured values
with standard solution
Perform a calibration
Check and document analyser
status
Perform a measurement and
check the values
if neces-
sary
Check viton tubes and replace if
necessary
Date: Signature: