04E3920 User Manual QuickTOC
®
purity
ATEX 2 109
7 How to Work With the Analyser
7.2 Configuration
LAR | PROCESS ANALYSERS AG
7.2.12.5 Switching the Condensate Pump On and Off
1. Press the "End maintenance" button.
2. Press the "Condensate pump off" button To switch on the condensate pump, press the
"Condensate pump on" button. When all maintenance has been completed, the "End
maintenance" button must be pressed. Only then can a measurement be started.
7.2.13 Single Measurement
With the single measurement single calibrations or samples can be measured. The single measurement
can be used to check calibrations.
Fig. 72: Single measurement screen
The current, the calibration path / sample path and the three parameters (repeat measurement, outlier
Notice
The heating time of the oven is about 120 minutes (2 hours). This corresponds to a
heating rate of 10° C / min. This time is fixed and not changeable.
Notice
The measured values of the single measurement are not saved in the database and
in the "Status screen" and are not transmitted to the analog outputs.
The measured values are only displayed in the "Single measurement" screen.
During the online measurement, the buttons "Select current" and "Start single mea-
surement" are deactivated. A single measurement can only be performed in an offline
mode.
i
i