3-66 SECTION 3 - OPERATING INSTRUCTIONS
Drift measurement
3:125. Program design recommendations for the drift (Doppler skymap) measurements are summarized in
Table 3-10
Table 3-10. Program Design Recommendations for Drift Measurements
Set CIT time to 20 (day) or
40 seconds (night) by using
appropriate inter-pulse peri-
od, number of pulses per fre-
quency, and frequency multi-
plexing.
Improved Doppler resolution for a
better signal analysis and echo reso-
lution in skymaps
Higher EMI on radio systems, long-
er minimum running time. Doppler
frequency bandwidth at 40 sec CIT
may not be sufficient for polar loca-
tions.
VALID SCHEDULES SUBOPTIMAL FOR SCIENCE OBSERVATIONS
3:126. The following deficiencies of schedule designs are not reported automatically:
• No loopback CCEQ program is included,
• No BIT program is included,
• Gain evaluation program uses constant gain setting different from subsequent science programs, and
• CCEQ or Gain evaluation program use incompatible frequency range and step with subsequent science
programs.