EasyManua.ls Logo

LDI DIGISONDE 4D - Tracker Card Calibration

Default Icon
321 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LDI Intellectual Property.
Not for secondary distribution or replication, in part or entirety.
DIGISONDE-4D
SYSTEM MANUAL
VERSION 1.2.11
6-8 SECTION 6 - MAINTENANCE
6:10. To remove the RF chassis foam filters:
Pull the RF Amp chassis fully forward or remove it entirely (see section 6:75). Locate the fan
filter on the rear panel. Unclip the plastic filter holder that clips over the edges of the fan. Re-
move the holder and take the filter out.
6:11. Fans - At least once a year all the fans in the system should be checked for normal quiet operation.
Case 2 fans
Upper chassis 3 fans
Data Computer 1 CPU fan
Control computer 1 CPU fan
Lower chassis 2 fans
Transmitter heat sink (2018 and later) 2 fans
28 volt power supply 1 fan
TRACKER CARD CALIBRATION
6:12. The following procedure describes the process of running and evaluating a tracker calibration. Tracker
cards mostly consist of several analog filters whose components are subject changes in performance due to ag-
ing. Performing a tracker calibration corrects for these changes.
6:13. Run Tracker Calibration and Cross Channel Equalization (CCEQ) programs.
6:14. Make sure DCART is not operating automatically by clicking the STOP button in the upper left corner
of the DCART screen.
6:15. Navigate to DCART / PROGSCHED tab (either EDITED or ACTIVE).
6:16. Select “Prog” on from the list of tabs on the left, under the PROGSCHED tab.
6:17. Locate and select a program labeled “Tracker Cal”, and click the “Run selected program
Figure 6-5: Tracker Cal Program
6:18. Locate and select a program labeled CCEQ” and click the Run selected program

Table of Contents