10-26
10.6.7 Arrhythmia Recall
Select ARR RECALL >> in the ARR ANALYSIS menu. The following menu appears. You can
review any stored arrhythmia event in this menu.
Figure 10-18 Arrhythmia Recall
You can perform the following operations:
UP-DOWN A maximum of 10 arrhythmia events can be displayed in the
window each time. In case of more than 10 events, you can
use the UP-DOWN option to review more. At most 8 pages
can be reviewed.
CURSOR This option allows you to select an arrhythmia event
displayed in the window.
DELETE This option allows you to delete a selected arrhythmia event.
RENAME This option allows you to change the name of a selected
arrhythmia event.
Select this option, and rotate the control knob till the desired
name appears. Then, press the control knob to select the
name.
WAVE >> Selecting this option opens the following window. In the
window, the waveform and the time of a selected arrhythmia
event as well as the parameter values at the event time are
displayed.