6-5
6.3.2 Alarms Paused
To pause all alarms of the monitor for 1, 2 or 3 minutes, press the SILENCE key on the control
panel once (for less than 2 seconds). In Alarms Paused status,
îš„ Visual alarms and audible alarms are both paused.
îš„ The parameters generating physiological alarms and their upper or lower limits stop
flashing.
îš„ Alarm messages are not displayed.
îš„ The physiological alarms area shows the rest time of alarms paused status.
îš„ The
icon will be displayed in the sound icon area.
The monitor will not terminate the Alarm Paused status even if a new technical or physiological
alarm occurs during the alarms paused time.
6.3.3 System Silenced
To silence the system, press the SILENCE key for no less than 2 seconds. In the System Silenced
status, all system sounds are shielded and the
icon is displayed in the sound icon area.
However, other modes of alarms (excluding audible alarms) are given as normal. If a new alarm
occurs, the System Silenced status will be terminated.
The system sounds include the audible alarms, key tones, heart beat tones and pulse tones.
Among them, key tones refer to the sounds produced when the control knob is rotated and
pressed.
6.3.4 Alarm Sound Off
In Alarm Sound Off status, all audible alarms are suppressed, but other modes of alarms and
other sounds are not influenced. The
icon will be displayed in the sound icon area. To turn
off the audible alarms, first access the USER MAINTAIN menu, and then set the ALARM
SOUND OFF to DISABLE. For details, please refer to 4.7 Maintenance.

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals