4-27
Trigger Conditions
A nurse call signal will be triggered only if all the following conditions are met:
1. The nurse call function is enabled.
2. An alarm of the preset alarm level and alarm type comes out.
3. The monitor is not in the Alarms Paused or the System Silenced status.
NOTE
z If no option in ALM LEV or ALM TYPE is selected, the nurse call signal will not
be triggered in whatever condition.
z The nurse call function can’t be used as the main alarm notice method. Medical
staff must combine the audible and visual alarms, the clinical vital signs of the
patient to determine the patient’s situation.
z In the Alarms Paused or the System Silenced status, the nurse call function of the
monitor will be disabled automatically.
4.7.5 CO
2
User Maintain
Selecting CO2 USER MAINTAIN >> in USER MAINTAIN menu opens the CO2 USER
MAINTAIN menu. The options contained in this menu are relative with the CO
2
module that you
monitor is equipped with. For details, please refer to chapter NIBP Monitoring.
4.7.6 IBP Pressure Calibration
The monitor provides the function of IBP pressure calibration.

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals