8-3
8.3 Recorder Operations
Continuous real-time recording
1. Press the RECORD key to start recording.
2. Press the RECORD key again to stop the recording.
8-second real-time recording
1. Press the RECORD key to start recording.
2. The recording stops automatically in 8 seconds.
Automatic recording
1. The recorder starts recording automatically at the preset interval (RT REC TIME).
2. The recording stops automatically in 8 seconds.
Alarm recording
1. When an alarm occurs, the recorder starts recording automatically.
2. The recording stops automatically when the preset alarm recording time (ALM REC TIME)
is over.
Frozen waveform recording
1. Press the FREEZE key to open the FROZEN menu.
2. Select WAVE 1, WAVE 2 and WAVE 3.
3. Select the REC option to record.
4. When the recording completes, the recorder stops automatically.
Trend graph recording
1. Select TREND GRAPH>> in SYSTEM MENU to open the TREND GRAPH window.
2. Select the REC option to start recording.
3. When the recording completes, the recorder stops automatically.
Trend table recording
1. Select TREND TABLE>> in SYSTEM MENU to open the TREND TABLE window.
2. Select the REC option to start recording.
3. When the recording completes, the recorder stops automatically.

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals