15-13
15.3.2 Preparations for CO
2
Measurement
1. Plug the sampling line into its receptacle before the measurement.
2. Open the CO2 SETUP menu and set WORK MODE to MEASURE.
3. The “CO2 START UP” prompt information is displayed on the screen until the startup is
finished.
Figure 15-7 CO
2
Module Airway Connections
When the monitor is powered on for the first time, the CO
2
module enters the STANDBY mode
by default. To activat the CO
2
module, you must change the work mode to MEASURE. The work
mode of the CO
2
module keeps unchanged when the monitor is restarted. For instance, if the
monitor is power off when the CO
2
module is working in the MEASURE mode, the CO
2
module
will automatically enter the MEASURE mode after the monitor is restarted. For details about
WORK MODE, please refer to the section 15.3.3CO2 .
WARNING
z Do not use the accessory if the packaging or the internal accessory is damaged.
Return it to the manufacturer.
z The sidestream sampling line is disposable. It should not be disinfected for reuse or
cross-used by different patients.