3.5 Powering on the Monitor
After installing the monitor, please follow the procedures described below to power on the
monitor:
1. Before using the monitor, please carry out corresponding safety inspection as given in 16.1
Inspection.
2. Press the Power Switch on the control panel. The alarm indicator will flash once in yellow
and red. Then, a beep will be heard.
3. The system begins self-testing and the product model will be displayed on the screen.
4. Several seconds later, the system finishes the self-test and displays the main screen.
5. The system will initiate every module, and display “XX alarm disabled!” information in the
lower left part of the screen. “XX” represents the name of every module, such as NIBP,
RESP etc.
6. At this time, you can operate the monitor using the control panel. “XX alarm disabled!”
information will disappear a few seconds later.
NOTE
z During initialization process, alarms of every module detected by the system are
useless, and thereby are disabled.