16-1
16 Maintenance
WARNING
z Failure on the part of the responsible hospital or institution employing the use of
the monitoring equipment to implement a satisfactory maintenance schedule
may cause undue equipment failure and possible health hazard.
z The safety inspection or maintenance, which requires opening the monitor
housing, must be performed by trained and authorize personnel only.
Otherwise, equipment failure and possible health hazard may be caused.
16.1 Inspection and Maintenance
Make sure the qualified service personnel have implemented a complete inspection before
putting the monitor into operation, after monitor servicing or system upgrading, or after the
monitor has been used for 6-12 consecutive months. This is to ensure the normal operation of the
system.
Follow these guidelines when inspecting the equipment:
The environment and the power supply meet the specified requirements.
Inspect the keys, control knob, connectors and accessories for damage.
Inspect the power cords for fraying or other damage and check the insulation.
The grounding cables are correctly connected.
Only specified accessories like electrodes, sensors and probes are applied.
The monitor clock is correct.
The audible and visual alarms functions normally.
The recorder functions normally and the recorder paper meets the requirement.
Batteries are in good status and can operate normally.
The defibrillator synchronization function must be verified according to your hospital regulations,
and be checked by a qualified technician once every 3 months.
In case of any damage or exception, do not use the monitor. Contact the technician in your
hospital or our Customer Service immediately.
CAUTION
z Be sure to check and maintain batteries regularly. For detailed battery check
and maintenance procedures and battery guidelines, see Section 2.5 Batteries.