4.7.1 Alarm Setup
Select ALARM SETUP>> to enter the following menu:
Figure 4-23 Alarm Setup
ALARM SOUND OFF
ENABLE: ALARM VOL can be set to 0.
DISABLE: ALARM VOL cannot be set to 0.
AUDIO OFF
MODE 1: the alarm sound interval of high level alarm is
8s and that of medium level alarm is 24s.
MODE 2: the alarm sound interval of high level alarm is
3s and that of medium level alarm is 14s.
PARA ALM TYPE LATCH, UNLATCH
REMINDER TONE Options: ON,OFF.
REMINDER INTERVAL Options: 1MIN, 2MIN and 3MIN.
ECG LEAD OFF Alarm level: HIGH, MED, LOW.
SpO
2
SENSOR OFF Alarm level: HIGH, MED, LOW.
WARNING
z Select the setting of ALARM SOUND OFF with caution.
NOTE
z In the case that alarm volume is set to 0, if you set ALARM SOUND OFF to
DISABLE, alarm volume change to the default.