A-13
A.11 NIBP Specifications
Measurement technique Auto oscillation
Displayed parameters Systolic pressure, diastolic pressure, mean pressure and PR
Mode of operation Manual, auto and continuous
Measurement interval in
auto mode
1/2/3/4/5/10/15/30/60/90/120/180/240/480 minutes
Measurement time in
continuous mode
5 minutes
mmHg Adult Pediatric Neonate
Systolic pressure 40 to 270 40 to 200 40 to 135
Diastolic pressure 10 to 210 10 to 150 10 to 100
Measurement range in
normal mode
Mean pressure 20 to 230 20 to 165 20 to 110
Measurement precision
Maximum average error: ±5mmHg
Maximum standard deviation: 8mmHg
Resolution 1mmHg
Static pressure
measurement range
0 to 300mmHg
Static accuracy ± 3 mmHg
Over-pressure protection
by software
Adult:
Pediatric:
Neonate:
297±3 mmHg
240±3 mmHg
147±3 mmHg
Over-pressure protection
by hardware
Adult:
Pediatric:
Neonate:
330 mmHg
330 mmHg
165 mmHg
Default start pressure
Adult:
Pediatric:
Neonate:
178±5 mmHg
133±10 mmHg
67±5 mmHg
PR from NIBP
Measurement range 40 to 240 bpm
Precision ±3 bpm or ±3%, whichever is greater
Resolution 1 bpm

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals