A-12
A.10.3 Nellcor SpO
2
Module
All SpO
2
sensors specified in the section Nellcor SpO2 Accessories meets the following
specifications when used with Nellcor SpO
2
module.
Sensor Range Precision*
MAX-A, MAX-AL, MAX-N,
MAX-P, MAX-I and MAX-FAST
70 to 100%
0% to 69%
±2%
Undefined
OxiCliq A, OxiCliq N, OxiCliq P,
OxiCliq I
70 to 100%
0% to 69%
±2.5%
Undefined
D-YS, DS-100A, OXI-A/N and
OXI-P/I
70 to 100%
0% to 69%
±3%
Undefined
SpO2 measurement range
and precision
MAX-R, D-YSE and D-YSPD
70 to 100%
0% to 69%
±3.5%
Undefined
PR measurement range and
precision
20 to 250 bpm: ±3 bpm
251 to 300 bpm: Undefined
Refreshing rate 1 s
Averaging time 8 s, 16 s
*: When sensors are used on neonatal subjects as recommended, the specified precision range is
increased by ±1%, to account for the theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood.