13-4
CONFIG >> You can select this option to access the TEMP CONFIG
menu, in which you may select FACTORY CONFIG or
USER CONFIG. After finishing your selection and exiting
the menu, the system pops up a dialog box asking for your
confirmation.
Temperature alarm limits:
Parameter Upper Limit Lower Limit Step
T1 and T2 (lower limit+1) ~50 0~(upper limit-1) 0.1
TD (lower limit+1) ~50 0~(upper limit-1) 0.1
Temperature alarm limits in factory configuration:
T1/T2 alarm high limit: 39℃
T1/T2 alarm low limit: 36℃
TD alarm limit: 2.0℃
13.4 Maintenance and Cleaning
WARNING
z Before cleaning the monitor or the probe, make sure the equipment is turned off
and disconnected from AC power.
Reusable Temperature Probes
The temperature probe should not be heated to a temperature over 100℃ (212℉). It is only
able to bear the temperature between 80 and 100℃ (176 to 212℉) for a short time.
The probe must not be disinfected in steam.
Only detergents containing alcohol can be used for disinfection.
The rectal probes should be used, if possible, in conjunction with a protective rubber cover.
To clean the probe, hold the tip with one hand and with the other hand rub the probe down
in the direction of the connector using a moist lint-free cloth.
NOTE
z Disposable temperature probes must not be re-sterilized or reused.
z Disposable temperature probes must be recycled or disposed of properly.