6-8
6.6 Testing Alarms
When the monitor starts up, a selftest will be performed automatically. The alarm indicator will
flash once in yellow and red. Then, the system gives a beep indicating that the visible and audible
alarm indicators are functioning correctly. The start-up screen is displayed.
For further testing of individual measurement alarms, perform the measurement on yourself (for
example SpO2 or CO2) or use a simulator. Adjust alarm limits and check if alarms behave
correctly.
6.7 When an Alarm Occurs
WARNING
z When an alarm occurs, always check the patient’s condition first.
When an alarm occurs to the monitor, refer to the following steps and take action properly.
1. Check the patient’s condition.
2. Identify the alarming parameter and the alarm category.
3. Identify the cause of the alarm.
4. Take action to remedy the alarm cause.
5. Check if the alarm is cleared.
For details about how to deal with specific alarms, see chapter Alarm Messages and Prompt
Information.