10-31
For different patient types, the upper/lower limits of the respiration rate alarm may vary in the
following range.
Patient type Upper Limit Lower Limit Increment (breath/min)
Adult (lower limit+2) ~120 0~(upper limit-2) 1
Neonate/pediatric (lower limit+2) ~150 0~(upper limit-2) 1
APNEA ALM Apnea alarm
Determines whether the patient’s cessation of breath is an
apnea event. Range: 10–40 seconds.
SWEEP Waveform speed
Options: 6.25, 12.5 and 25.0 mm/s.
HOLD TYPE Calculation type
Options: AUTO and MANUAL
1. AUTO: When AUTO is selected, the monitor automatically determines the detection
threshold for respiration and calculates the respiration rate. The HOLD HI and HOLD LO
options are inactive.
2. MANUAL: When MANUAL is selected, the user determines the detection threshold for
respiration and the monitor calculates the respiration rate depending on the user-selected
criteria.
CONFIG >> Select CONFIG >> to access the RESP CONFIG menu. You
can select either FACTORY CONF or USER CONF. After
finishing the selection and exiting the menu, a dialog box
pops up asking for confirmation of your selection.